E. Schafer & D. Lohmann
Department of Operative Dentistry, University of Munster, Munster, Germany.
Aim.
To determine the cleaning effectiveness and the shaping ability of FlexMaster nickel-titanium rotary instruments and stainless steel hand K-Flexoашles during the preparation of curved root canals in extracted human teeth.
Conclusions.
With in the parameters of this study, manual instrumentation using K-Flexofiles resulted in better canal cleaning
than with rotary nickel-titanium FlexMaster instruments. FlexMaster instruments maintained the original curvature significantly better.
E. Schafer & D. Lohmann
Department of Operative Dentistry, University of Munster, Munster, Germany.
Introduction.
The quality guidelines of the European Society of Endodontology (1994) state that the elimination of residual pulp tissue, the removal of debris and the maintenance of the original canal curvature during enlargement are the main objectives of root-canal instrumentation. Scanning electronic microscopy is useful in examining root-canal cleanliness after preparation with different techniques or instruments. Several studies have concluded that none of the instrumentation techniques or devices completely clean root canals, especially when curved (Bolanos&Jensen1980, Haikel&Allemann1988, Hulsmann et al.1997).
Most of these authors also indicate that the cleaning ability of manual root-canal instrumentation is superior to automated devices (Mizrahi et al. 1975, Schwarze & Geurtsen 1996, Hulsmann et al. 1997). However, it was recently shown by several investigations that automated devices using rotary nickel-titanium instruments with various tapers led to good instrumentation results, even in severely curved root canals (Thompson & Dummer 1997, Kum et al. 2000). Unfortunately, little is known about the cleaning effectiveness of these systems, but most recently it has been observed that the different rotary nickel-titanium instruments showed inconsistent results (Hulsmann et al. 2000, Schafer & Zapke 2000).
The aim of this investigation was to compare the cleaning efficacy (residual debris, quality of the smear layer) after preparation of severely curved root canals with rotary nickel-titanium FlexMaster (VDW, Antaeos, Munich, Germany) or with stainless steel hand K-Flexofiles (Dentsply Maillefer, Ballaigues, Switzerland).Moreover, another purpose of this study was to assess whether instrumentation had an effect on canal curvature.
Materials and methods.
Selection of teeth.
A total of 48 extracted human teeth with at least one curved root and curved root canal were selected for this investigation. Coronal access was achieved using diamond burs. Only teeth whose clinical crowns were largely intact, whose root canals were freely accessible with a root-canal instrument size10 up to the intact root tip, and whose root-canal width near the apex were approximately compatible with size 15 were included. This was checked with silver points sizes 15 and 20 (VDW, Antaeos, Munich, Germany).
Standardized radiographs were taken prior to instrumentation with the initial root-canal instrument of size 15 inserted into the curved canal. The tooth was attached to a Kodak Ultra-speed ¢lm (Kodak, Stuttgart, Germany) and was aligned so that the long axis of the root canal was parallel, and as near as possible to the surface of the ¢lm. The X-ray tube, and thus the central Xray beam, was aligned perpendicular to the root canal. The exposure time (0.12 s;70 kV,7 mA) was same for all radiographs with a constant source-to-¢lm distance of 50 cm and an object-to-¢lm distance of 5 mm. The ¢l ms were developed, ¢fixed, and dried in an automatic processor (Durr-Dental XR 24 Nova, Durr, Bietigheim-Bissingen, Germany).
Table 1. Characteristics of curved root canals (n= 24 teeth per group).
* n= 24 canals in each group.
The degree and the radius of canal curvature were determined using a computerized digital image processing system(Schafer et al. in press).Only, teeth whose radii of curvature ranged between 4 and 9 mm and whose angles of curvature ranged between 258 and 358 were included (Table 1). On the basis of the degree and the radius of curvature, the teeth were allocated into two identical groups of 24 teeth. The homogeneity of the two groups with respect to the degree and the radius of curvature was assessed using a t-test (Table 1). At the end of canal preparation, the canal curvatures were re determined on the basis of a radiograph with the ¢final root-canal instrument inserted into the canal using the same technique (Schafer et al. in press) in order to compare the initial curvatures with those after instrumentation. Only one canal was instrumented in each tooth.
Root-canal instrumentation.
The working length was obtained by measuring the length of the initial instrument (size10) at the apical foramen minus 1mm for all groups. The canals of all teeth were prepared with instruments up to size 35; instruments were used to enlarge two canals only. After each instrument, the root canal was £flushed with 5 mL of a 2.5% NaOCl solution and at the end of instrumentation with 5 mL of NaCl using a plastic syringe with a closed-end needle.
The following instrumentation sequences were used with the two instruments:
Group A
FlexMaster: these instruments were set into permanent rotation (250 rpm) with a 8 :1 reduction handpiece (Type 5059, Nouvag, Goldach, Switzerland) powered by a torque-limited electric motor (TCM Endo 2, Nouvag, Konstanz, Germany) using torque setting 2,which is stated to be equivalent to a torque limitation of 1.5- 1.7 Ncm by the manufacturer. Instrumentation was completed using the crown-down technique according to the manufacturer’s instructions using a gentle in-and-out motion. Every instrument was withdrawn when resistance was felt and replaced by the next (smaller) instrument in the sequence. The preparation sequence was the same as described in part 1 of this two-part report (Schafer & Lohmann 2002):
Once, the instrument had negotiated to the end of the canal and had rotated freely, it was removed.
Group B
K-Flexofile: hand instrumentation with these stainless steel instruments with non-cutting tips was performed using a reaming motion. All canals were sequentially prepared from size15 up to 35 without pre-bending the instruments, which were used to the full working length.
Evaluations.
All root-canal preparations were completed by one operator, whilst the scanning electron microscope (SEM) evaluations and the assessment of the canal curvatures prior to and after instrumentation were carried out by a second examiner who was blind with respect of all to the experimental groups.
Canal cleanliness.
After preparation, all root canals were £flushed with sodium chloride and dried with adsorbent paper points. Roots were split longitudinally, prepared for SEM investigation and examined under the SEM (Philips PSEM 500_, Eindhoven, the Netherlands) at 20-2500_ magnification. Separate evaluations were recorded for debris and smear layer. The cleanliness of each root canal was evaluated in three areas (apical, middle, and coronal third of the root) by means of a numerical evaluation scale (Hulsmann et al.1997). The following scheme was used: Debris (dentine chips, pulp remnants, and particles loosely attached to the canal wall):
Smear layer (dentine particles, remnants of vital or necrotic pulp tissue, bacterial components, and retained irrigant):
The data established for scoring the debris and the smear layer were separately recorded and analyzed statistically. Owing to the ordinal nature of the scores, the data were subjected to the Wilcoxon test (P < 0.05).
Instrumentation results.
Based on the canal curvatures assessed prior to and after the instrumentation, canal strength was determined as the difference between canal curvature prior to and after the instrumentation. The t-test was used for comparison of the two groups. The level of statistical significance was set at P<0.05.
The time for canal preparation was recorded, and the total active instrumentation, instrument changes within the sequence and irrigation was included. The change of working length was determined by subtracting the ¢final length (measured to the nearest 0.5 mm) of each canal after preparation from the original length. The preparation time and the change of working length were analyzed statistically using the t-test (preparation time) and the Mann-Whitney U-test (change of working distance) at a significance level of P<0.05. The number of fractured and permanently deformed instruments during the enlargement was also recorded.
Table 2. Summary of scores for debris.
* Listed are the number of canal areas evaluated as scores 1-5 (n= 24 teeth per group). Three canal areas (coronal, middle and apical thirds) have been evaluated per tooth, thus resulting in a total of 72 canal areas per group. Score1indicates the best and score 5 theworst canal cleanliness.
During the preparation of the 48 canals, no instruments separated, and only one FlexMaster and one K-Flexofile permanently deformed.
Canal cleanliness.
The scores for the debris and the smear layer are detailed in Tables 2 and 3. Completely cleaned root canals were not found with any of the two instruments. On average, more effective cleaning was observed in the coronal and the middle thirds of the canals (Fig. 1).
In general, the use of K-Flexofiles resulted in significantly less debris (P < 0.001) (Fig. 2) and less smear layer (P < 0.05) compared to the canal preparation with Flex-Master instruments (Tables 2 and 3); these differences were not significant in the apical third.
Figure 1. Canal wall after preparation with FlexMaster rotary nickel-titanium instruments.
(a) Clean canal wall with only very small debris particles in the middle portion of the prepared canal (score1, original magnification 40x).
(b) Apical portion of the canal: Complete or nearly complete covering of the canal wall by debris after preparation (score 5, original magnification 80x).
Table 3. Summary of scores for smear layer.
*The number of canal areas evaluated as scores 1-5 (n=24 teeth per group).Three canal areas (coronal, middle and apical thirds) have been evaluated per tooth, thus resulting in a total of 72 canal areas per group. Score1indicates the best and score 5 the worst result.
Figure 2. Clean surface in the apical third of a curved root canal after manual instrumentation using K-Flexofiles in a reaming working motion (score1, original magnification 40x).
Figure 3. Straightening of the curved canals after preparation with the two different instruments (n=24 canals in each group): Combined box-and-whisker, and dot plot, each dot represents a reading of the difference between canal curvature prior to and after instrumentation.
Table 4. Mean preparation time and SD, and mean changes of working distance and SD in the two different instruments.
Table 5. Mean degree of straightening of curved canals (8) and SD after canal preparation in the two different instruments.
* n= 24 canals in each group.
References.
American Associationof Endodontists (1998) Glossary. Contemporary Terminology for Endodontics, 6th edn.
Bechelli C, Zecchi Orlandini S, Colafranceschi M (1999) Scanning electron microscope study on the efficacy of root canal wall debridement of hand versus Light speed instrumentation. International Endodontic Journal 32, 484-93.
Bolanos OR, Jensen JR (1980) Scanning electron microscope comparisons of the efficacy of various methods of root canal preparation. Journal of Endodontics 6, 815-22.
European Society of Endodontology (1994) Consensus report of the European Society of Endodontology on quality guidelines for endodontic treatment. International Endodontic Journal 27,115-24.
Gambarini G (1999) Shaping and cleaning the root canal system: a scanning electron microscopic evaluation of a new instrumentation and irrigationtechnique. Journal of Endodontics 25,800-3.
Haikel Y, Allemann C (1988) Effectiveness of four methods for preparing root canals: a scanning electron microscopic evaluation. Journal of Endodontics14, 340-5.
H lsmann M (2000) Entwicklung Einer Methodik Zur Standardisierten Uberprufung Verschiedener Aufbereitungsparameter und Vergleichende in-Vitro-Untersuchung Unterschiedlicher Systeme Zur Maschinellen Wurzelkanalaufbereitung (Habilitationsschrift). Gottingen, Germany: University of Gottingen. Hulsmann M, Rummelin C, Schafers F (1997) Root canal cleanliness after preparation with different endodontic handpieces and hand instruments: a comparative SEM investigation. Journal of Endodontics 23,301-6.
HulsmannM, Versumer J, SchadeM(2000) Acomparative study of Lightspeed, ProFile 0.04, Quantec and Hero 642. International Endodontic Journal 33,150 (abstract).
Kum K-Y, SpOngberg L, Cha BY, Il-Young J, Seung-Jong L, Chan- Young L (2000) Shaping ability of three ProFile rotary instrumentation techniques in simulated resin root canals. Journal of Endodontics 26,719-23. Mizrahi SJ, Tucker JW, Seltzer S (1975) A scanning electron microscopic study of the efficacy of various endodontic instruments. Journal of Endodontics1,324-33.
Nair PNR, Sjogren U, Krey G, Kahnberg KE, Sundqvist G (1990) Intraradicular bacteria and fungi in root-filled, asymptomatic human teeth with therapy-resistant periapical lesions: a long-term light and electron microscopic follow-up study. Journal of Endodontics16,580-8.
PetscheltA, Stumpf B, RaabW(1987) Tightness of root canal sealers with and without smear layer. Deutsche Zahnarztliche Zeitschrift 42,743-6.
Schafer E, Diez C, Hoppe W, Tepel J (in press) Roentgenographic investigation of frequency and degree of canal curvatures in human permanent teeth. Journal of Endodontics in press.
Schafer E, Lohmann D (2002) Efficiency of rotary nickeltitanium FlexMaster instruments compared with stainless steel hand K-Flexofile. Part 1. Shaping ability in simulated curved canals. International Endodontic Journal 35, 514-21.
Schafer E, Zapke K (2000) A comparative scanning electron microscopic investigation of the efficacy of manual and automated instrumentation of root canals. Journal of Endodontics 26, 660-4.
Schwarze T, GeurtsenW (1996) Comparative qualitative SEM study of automated vs. hand instrumentation of root canals. Deutsche Zahnarztliche Zeitschrift 51, 227-30.
SpOngberg L, Engstrom B, Langeland K (1973) Biological effects of dental materials. Part 3. Toxicity and antimicrobial effects on endodontic antiseptics in vitro. Oral Surgery 36, 856-71.
Thompson SA, Dummer PMH (1997) Shaping ability of ProFile .04 taper Series 29 rotary nickel-titanium instruments in simulated canals. Part 1 and 2. International Endodontic Journal 30,1-15.
Turkun M, Cengiz T (1997) The effects of sodium hypochlorite and calcium hydroxide on tissue dissolution and root canal cleanliness. International Endodontic Journal 30, 335-42.
West JD, Roane JB, Goerig AC (1994) Cleaning and shaping the root canal system. In: Cohen S, Burns RC, eds. Pathways of the Pulps,6th edn. St. Louis, MO, USA: Mosby.179-218.
Wu MK,WesselinkPR (1995) Efficacy of three techniques in cleaning the apical portion of curved root canals. OralSurgery 79, 492-6.
Yamada RS, Armas A, Goldman M, Lin PS (1983) A scanning electron microscopic comparison of high volume final flush with several irrigating solutions. Part 3. Journal of Endodontics 9, 137-42.